The 50™ TS Washer offers functionality that is unsurpassed in its class. The color touchscreen provides a visual interface that makes creating protocols fast and intuitive. Its performance for conventional plate washing is excellent, and it is also ideal for cell-based assay washing and for processing biomagnetic or polystyrene bead and vacuum filtration protocols. Used in conjunction with the 800™ TS Absorbance Reader or other detection system, the 50 TS brings convenience and consistently high quality results to your laboratory’s plate washing workflows.
Agilent BioTek 50 TS Washer
The 50™ TS Washer offers functionality that is unsurpassed in its class. The color touchscreen provides a visual interface that makes creating protocols fast and intuitive. Its performance for conventional plate washing is excellent, and it is also ideal for cell-based assay washing and for processing biomagnetic or polystyrene bead and vacuum filtration protocols. Used in conjunction with the 800™ TS Absorbance Reader or other detection system, the 50 TS brings convenience and consistently high quality results to your laboratory’s plate washing workflows.
General
Microplate types | 24-, 96-, 384-well plates (96- and 384- well: standard and low profile plates; solid and filter bottom (option), 24-well plates with 4-well manifold) |
Onboard software | Up to 75 user-programmable protocols Quick menu Create or edit custom protocols Run protocols created onboard or downloaded from LHC™ Software |
Software (computer control) | Optional: Liquid Handling Control™ (LHC), for external computer control and operation LHC Software LHC Secure for 21 CFR Part 11 compliance |
Separation | Biomagnetic separation (“M” configurations) Vacuum filtration (“F” configurations) |
Shaking | Programmable in minutes and seconds up to 30 minutes 5 intensities from 15-19 Hz |
Soaking | Programmable in minutes and seconds up to 30 minutes |
User interface | 4.3" color LCD touchscreen display |
Microplate types | 24-, 96-, 384-well plates (96- and 384- well: standard and low profile plates; solid and filter bottom (option), 24-well plates with 4-well manifold) |
Onboard software | Up to 75 user-programmable protocols Quick menu Create or edit custom protocols Run protocols created onboard or downloaded from LHC™ Software |
Software (computer control) | Optional: Liquid Handling Control™ (LHC), for external computer control and operation LHC Software LHC Secure for 21 CFR Part 11 compliance |
Separation | Biomagnetic separation (“M” configurations) Vacuum filtration (“F” configurations) |
Shaking | Programmable in minutes and seconds up to 30 minutes 5 intensities from 15-19 Hz |
Soaking | Programmable in minutes and seconds up to 30 minutes |
User interface | 4.3" color LCD touchscreen display |
Washing
Manifold types | 96-well washing: 8-well, 8-well short tube, 2x8-well, 12-well manifolds 96-384-well washing: 16-well manifold 24-well washing: 4-well manifold |
Volume range | 25-3,000 µL-well |
Wash cycles | 1-10 |
Buffer/reagent selection | Auto switching module for up to 3 buffers ("V" configurations) |
Supply bottle | 2L 4L (optional) |
Dispense precision | 96-well plate: 8- and 8s-well manifold, 300 µL/well: <3.0% CV 12-well manifold, 300 µL/well: <3.0% CV 2x8-well manifold, 300 µL/well: <4.0% CV 384-well plate, 8-, 16-well manifold, 100 µL/well: <4.0% CV 24-well plate, 4-well manifold, 120 µL/well): <4.0% CV |
Residual volume | 96-well plate: 8-, 8s-, 12-well manifold,3 cycles, 300 µL/well: <2.0 µL/ well 2x8-well manifold, 3 cycles, 300 µL/well: <4.0 µL/ well Filter bottom, 8-, 2x8-, 12-well manifold: <1.2 g increase after blotting 384-well plate: 8-, 16-well manifold, 1 cycle, 100 µL/ well: <4.0 µL/ well 24-well plate: 4-well manifold, 1120 µL/ well: <50 µL/well |
Wash speed | 96 wells: 2x8 manifold, 3 cycles, 300 µL/well: <80s 12-well manifold, 3 cycles, 300 µL/well: <90s 8- and 8s-well manifold, 3 cycles, 300 µL/well: <130s 384 wells, 8- and 16-well manifold, 3 cycles, 100 µL/well: <260s 24 wells, 4-well manifold,1 cycle, 1120 µL/well: <60s |
Wash volume | 25-3,000 µL well |
Waste bottle | 2L 4L (optional) |
Flow rates | 150 to 1000 µL/well/second (manifold dependent) |
Sterilization | Chemical |
Fluid delivery | One positive displacement syringe drive |
Manifold types | 96-well washing: 8-well, 8-well short tube, 2x8-well, 12-well manifolds 96-384-well washing: 16-well manifold 24-well washing: 4-well manifold |
Volume range | 25-3,000 µL-well |
Wash cycles | 1-10 |
Buffer/reagent selection | Auto switching module for up to 3 buffers ("V" configurations) |
Supply bottle | 2L 4L (optional) |
Dispense precision | 96-well plate: 8- and 8s-well manifold, 300 µL/well: <3.0% CV 12-well manifold, 300 µL/well: <3.0% CV 2x8-well manifold, 300 µL/well: <4.0% CV 384-well plate, 8-, 16-well manifold, 100 µL/well: <4.0% CV 24-well plate, 4-well manifold, 120 µL/well): <4.0% CV |
Residual volume | 96-well plate: 8-, 8s-, 12-well manifold,3 cycles, 300 µL/well: <2.0 µL/ well 2x8-well manifold, 3 cycles, 300 µL/well: <4.0 µL/ well Filter bottom, 8-, 2x8-, 12-well manifold: <1.2 g increase after blotting 384-well plate: 8-, 16-well manifold, 1 cycle, 100 µL/ well: <4.0 µL/ well 24-well plate: 4-well manifold, 1120 µL/ well: <50 µL/well |
Wash speed | 96 wells: 2x8 manifold, 3 cycles, 300 µL/well: <80s 12-well manifold, 3 cycles, 300 µL/well: <90s 8- and 8s-well manifold, 3 cycles, 300 µL/well: <130s 384 wells, 8- and 16-well manifold, 3 cycles, 100 µL/well: <260s 24 wells, 4-well manifold,1 cycle, 1120 µL/well: <60s |
Wash volume | 25-3,000 µL well |
Waste bottle | 2L 4L (optional) |
Flow rates | 150 to 1000 µL/well/second (manifold dependent) |
Sterilization | Chemical |
Fluid delivery | One positive displacement syringe drive |
Physical Characteristics
Power | External 24VDC power supply compatible with 100-240VAC @ 50-60Hz |
Dimensions | 15 x 15 x 8 inches (35.6 x 40.6 x 16.5 cm) |
Weight | 22 lbs (9.8 kg) |
Connectivity | 1 USB port for computer control |
Power | External 24VDC power supply compatible with 100-240VAC @ 50-60Hz |
Dimensions | 15 x 15 x 8 inches (35.6 x 40.6 x 16.5 cm) |
Weight | 22 lbs (9.8 kg) |
Connectivity | 1 USB port for computer control |
Regulatory
Regulatory | CE and TUV marked. ROHS compliant. In Vitro Diagnostic use models are available. |
Configurations
Part #
|
96-well only | 96-/384-well | Buffer Switching | Biomagnetic Separation* | Vacuum Filtration |
---|---|---|---|---|---|
50TS8
|
•
|
||||
50TS8V
|
•
|
•
|
|||
50TS8M
|
•
|
•
|
|||
50TS8MV
|
•
|
•
|
•
|
||
50TS8F
|
•
|
•
|
|||
50TS8MF
|
•
|
•
|
•
|
||
50TS12
|
•
|
||||
50TS12V
|
•
|
•
|
|||
50TS16
|
•
|
||||
50TS16V
|
•
|
•
|
* With available accessory module sold separately